Calibration Toolkit

Connect via Serial Port, available on Chrome desktop version

Serial: Waiting for connection

Message Received: 0


Configuration Menu | Select a tab

Tools



Dissolved Oxygen | Docs


- I,Type,FirmwareVer


- CAL,0: None | CAL,1: One point | CAL,2: Two point

- No need for DO calibration (Cleared after reboot)

- Expose to air until stablilze
- Put into DO 0 liquid until stablilze


Potential of Hydrogen | Docs


- I,Type,FirmwareVer


- CAL,0: None | CAL,1: One point | CAL,2: Two point | CAL,3: Three point

- Set before calibration (Cleared after reboot)

- Put into pH7 until stablilze
- Put into pH4 until stablilze
- Put into pH10 until stablilze


Electrical Conductivity | Docs


- I,Type,FirmwareVer


- CAL,0: None | CAL,1: One point | CAL,2: Two point

- Set before calibration (Cleared after reboot)

  • - Set before calibration begin
    - Expose to air until stablilze
    - Put into 84uS until stablilze
    - Put into 1413uS until stablilze
  • - Set before calibration begin
    - Expose to air until stablilze
    - Put into 12880uS until stablilze
    - Put into 80000uS until stablilze
  • - Set before calibration begin
    - Expose to air until stablilze
    - Put into 12880uS until stablilze
    - Put into 150000uS until stablilze