Calibration Toolkit
Connect via Serial Port, available on Chrome desktop version
Connect
Serial:
Waiting for connection
Message Received:
0
Configuration Menu | Select a tab
Tools
DO Sensor
pH Sensor
EC Sensor
Tools
Exit Config Mode
Dissolved Oxygen |
Docs
Circuit Info
- I,Type,FirmwareVer
Read x5
Read x10
View Calibration info
- CAL,0: None | CAL,1: One point | CAL,2: Two point
Clear Calibration Data
Set Temperature Compensation
- No need for DO calibration (Cleared after reboot)
Calibration Air
- Expose to air until stablilze
Calibration Zero DO
- Put into DO 0 liquid until stablilze
Send Command Directly
Potential of Hydrogen |
Docs
Circuit Info
- I,Type,FirmwareVer
Read x5
Read x10
View Calibration info
- CAL,0: None | CAL,1: One point | CAL,2: Two point | CAL,3: Three point
Clear Calibration Data
Set Temperature Compensation
- Set before calibration (Cleared after reboot)
Calibration Mid Point
- Put into pH7 until stablilze
Calibration Low Point
- Put into pH4 until stablilze
Calibration High Point
- Put into pH10 until stablilze
Send Command Directly
Electrical Conductivity |
Docs
Circuit Info
- I,Type,FirmwareVer
Read x5
Read x10
View Calibration info
- CAL,0: None | CAL,1: One point | CAL,2: Two point
Clear Calibration Data
Set Temperature Compensation
- Set before calibration (Cleared after reboot)
K = 0.1
Calibration Set K to 0.1
- Set before calibration begin
Calibration Dry
- Expose to air until stablilze
Calibration Low Point
- Put into 84uS until stablilze
Calibration High Point
- Put into 1413uS until stablilze
K = 1.0
Calibration Set K to 1.0
- Set before calibration begin
Calibration Dry
- Expose to air until stablilze
Calibration Low Point
- Put into 12880uS until stablilze
Calibration High Point
- Put into 80000uS until stablilze
K = 10
Calibration Set K to 10
- Set before calibration begin
Calibration Dry
- Expose to air until stablilze
Calibration Low Point
- Put into 12880uS until stablilze
Calibration High Point
- Put into 150000uS until stablilze
Send Command Directly